Starting Download..
save Save

How Does 4-wire RTD Connect to 34970A?

Frequently Asked Questions (FAQs)

Summary

If you want to use 34970A/34972A/ DAQ970 and 4-wire RTD to measure temperature, you need to change the configuration settings to 4-wire RTD. Connect one RTD compensation wire at position ① and try to make the four wires have the same length. Then you can configure the measurement type as 4-wire RTD temperature and get temperature results with 34901A or 34902A.

Question

How Does 4-wire RTD Connect to 34970A?

Answer

If you want to use 34970A/34972A/ DAQ970 and 4-wire RTD to measure temperature, you need to change the configuration settings to 4-wire RTD as the picture shows above. Connect one RTD compensation wire at position ① and try to make the four wires have the same length. Then you can configure the measurement type as 4-wire RTD temperature and get temperature results with 34901A or 34902A.

image

See Also

Keysight link page 35
Was this helpful?


Didn't find what you're looking for?